Product
Philips Intellivue Patient Monitor Model MP70 (M8007A)
- FDA product code
- MHX — Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K093268
- Affected lot / code info
- Serial Numbers: DE22702728 - DE84399581
Why it was recalled
Additional Instruction for Use for Intellivue MP60/MP70 to prevent monitors from becoming loose and detaching from the mounting arm.
Root cause (FDA determination)
Labeling design
Action the firm took
Philips issued on 6/28/10 an Urgent Medical Device Correction notification/Field Safety Notice. Included with the Urgent Medical Device Correction notification/Field Safety Notice is an updated instructions for use and service instructions for these monitors that remind users to follow the mounting arm manufacturer's instructions for use when repositioning the monitor and instructing them to perform a periodic check that the monitor is not loose on the arm.
Recalling firm
- Firm
- Philips Healthcare Inc.
- Address
- 3000 Minuteman Road, Andover, Massachusetts 01810
Distribution
- Distribution pattern
- Worldwide distribution: USA, Canada, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, Bosnia, BRAZIL, BULGARIA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRI, LITHUANIA, LUXEMBOURG, MACAU, MALAYSIA, MAURITIUS, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA & MONTE NEGRO, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, United Arab Emirates, UNITED KINGDOM, UZBEKISTAN, VENEZUELA, VIETNAM, VIRGIN ISLANDS (BR), and YEMEN.
Timeline
- Recall initiated
- 2010-06-28
- Posted by FDA
- 2010-08-10
- Terminated
- 2017-02-15
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #92622. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.