FDA Device Recalls

Recalls /

#93261

Product

SCORPIO PS TIBIAL TRIAL, non sterile, made in the USA Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. The Tibial trial are utilized during surgery to assess the overall component fit, ligament stability, and joint range of motion prior to final implantation of the insert to the tray.

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
T72-3-0108M SCORPIO PS TIBIAL TRIAL, SIZE 1, 8 MM T72-3-0110M SCORPIO PS TIBIAL TRIAL, SIZE 1, 10 MM T72-3-0112M SCORPIO PS TIBIAL TRIAL, SIZE 1, 12 MM T72-3-0115M SCORPIO PS TIBIAL TRIAL, SIZE 1, 15 MM T72-3-0118M SCORPIO PS TIBIAL TRIAL, SIZE 1, 18 MM T72-3-0121M SCORPIO PS TIBIAL TRIAL, SIZE 1, 21 MM T72-3-0124M SCORPIO PS TIBIAL TRIAL, SIZE 1, 24 MM T72-3-0308M SCORPIO PS TIBIAL TRIAL, SIZE 3, 8 MM T72-3-0310M SCORPIO PS TIBIAL TRIAL, SIZE 3, 10 MM T72-3-0312M SCORPIO PS TIBIAL TRIAL, SIZE 3, 12 MM T72-3-0315M SCORPIO PS TIBIAL TRIAL, SIZE 3, 15 MM T72-3-0318M SCORPIO PS TIBIAL TRIAL, SIZE 3, 18 MM T72-3-0321M SCORPIO PS TIBIAL TRIAL, SIZE 3, 21 MM T72-3-0324M SCORPIO PS TIBIAL TRIAL, SIZE 3, 24 MM T72-3-0508M SCORPIO PS TIBIAL TRIAL, SIZE 5, 8 MM T72-3-0510M SCORPIO PS TIBIAL TRIAL, SIZE 5, 10 MM T72-3-0512M SCORPIO PS TIBIAL TRIAL, SIZE 5, 12 MM T72-3-0515M SCORPIO PS TIBIAL TRIAL, SIZE 5, 15 MM T72-3-0518M SCORPIO PS TIBIAL TRIAL, SIZE 5, 18 MM T72-3-0521M SCORPIO PS TIBIAL TRIAL, SIZE 5, 21 MM T72-3-0524M SCORPIO PS TIBIAL TRIAL, SIZE 5, 24 MM T72-3-0708M SCORPIO PS TIBIAL TRIAL, SIZE 7, 8 MM T72-3-0710M SCORPIO PS TIBIAL TRIAL, SIZE 7, 10 MM T72-3-0712M SCORPIO PS TIBIAL TRIAL, SIZE 7, 12 MM T72-3-0715M SCORPIO PS TIBIAL TRIAL, SIZE 7, 15 MM T72-3-0718M SCORPIO PS TIBIAL TRIAL, SIZE 7, 18 MM T72-3-0721M SCORPIO PS TIBIAL TRIAL, SIZE 7, 21 MM T72-3-0724M SCORPIO PS TIBIAL TRIAL, SIZE 7, 24 MM T72-3-0908M SCORPIO PS TIBIAL TRIAL, SIZE 9, 8 MM T72-3-0910M SCORPIO PS TIBIAL TRIAL, SIZE 9, 10 MM T72-3-0912M SCORPIO PS TIBIAL TRIAL, SIZE 9, 12 MM T72-3-0915M SCORPIO PS TIBIAL TRIAL, SIZE 9, 15 MM T72-3-0918M SCORPIO PS TIBIAL TRIAL, SIZE 9, 18 MM T72-3-0921M SCORPIO PS TIBIAL TRIAL, SIZE 9, 21 MM T72-3-0924M SCORPIO PS TIBIAL TRIAL, SIZE 9, 24 MM All lots.

Why it was recalled

Stryker Orthopaedics became aware that there is the potential for the Scorpio T72 tibial Insert Trials to be oversized by a maximum of .024" (0.6mm).

Root cause (FDA determination)

Other

Action the firm took

Stryker Howmedica Osteonics Corp.sent Urgent Product Correction letters Correction letters dated July 23, 2010, via Federal Express to all affected customers. were sent to all Stryker Branches on July 23, 2010 via Federal Express. Asia Pacific, Canada, Europe, Middle East, Africa, Japan and Latin America sites were sent letters on July 26, 2010 by Federal Express. The letters identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the attached Product Correction Acknowledgement Form by fax as soon as possible to 201-831-6069. For any questions customers were to call 201-972-2100 or 201-831-5028.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Drive, Mahwah, New Jersey 07430

Distribution

Distribution pattern
Worldwide Distribution - USA including OH, TX, MS, and CO and the countries of Australia, Hong Kong, Taiwan, India, Korea, Japan, Chile, Brazil, Columbia, Venezuela, , Panama, Santo Domingo, Mexico, Argentina, Costa Rica, Guatemala, Peru,Sweden, Germany, Netherlands, France, Spain, Italy, Romania, South Africa, United Kingdom. and Canada.

Timeline

Recall initiated
2010-07-13
Posted by FDA
2011-03-16
Terminated
2017-09-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #93261. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.