FDA Device Recalls

Recalls /

#93719

Product

Omnifit M/S PSL Porous Coated Acetabular Shell 50MM; Stryker Orthopaedics, Uses in primary and revision total hip arthroplasty which are indicated for painful, disabling joint disease resulting from degenerative arthritis, rheumatoid arthritis, post traumatic arthritis or late stage avascular necrosis.

FDA product code
MEHProsthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K912654
Affected lot / code info
Catalog number 2017-0050, Lot # 5L3074, 5T3024,5T30224C, 5V3057, 5V3057A, 7A3023, 7A3023C,7E3009C7K3007, 7K3020, 7S3014, 8K3107, 8K3107A, 8M3010, 8M3010A, 8M3010C, 8M3010D, 8M3010F, 8N3030, 8N3050A, 8N3050D, MET81K, MHNNM9, MHREMA.

Why it was recalled

Stryker Orthopaedics determined that specific lots of the Omnifit Acetabular Shells may have a screw hole wall thickness that falls above the design specification.

Root cause (FDA determination)

Device Design

Action the firm took

The firm, Stryker Orthopaedics, sent an "URGENT PRODUCT RECALL" letter dated August 2, 2010, via Federal Express to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory to identify the affected product; retrieve all affected product and return it their branch or agency warehouse for reconciliation; reconcile all product(s) on the attached Product Recall Acknowledgement Form and fax to (201) 831-6069, and return the affected product via Inter-Org Transfer and ship to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. If you have any questions, feel free to contact the Manager, Divisional Regulatory Compliance at (201) 972-2100.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Drive, Mahwah, New Jersey 07430

Distribution

Distribution pattern
Worldwide distribution: USA including states of: AL, AR, CA, FL, IN, ME, MI, MS, NC, NJ, OK, OR, WA , and WI; and countries including: China, Hong Kong and Korea.

Timeline

Recall initiated
2010-08-02
Posted by FDA
2011-02-07
Terminated
2012-11-09
Status

Source: openFDA Device Recall endpoint. Recall record ID #93719. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.