Recalls / —
—#94452
Product
AXIOM Artis Systems VB23A-VB23H with Perivision functionality, Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755 Angiographic X-ray System
- FDA product code
- IZI — System, X-Ray, Angiographic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K021021
- Affected lot / code info
- Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755. Mulitple Serial Numbers per Distribution Listing
Why it was recalled
Sporadic system crash while Perivision is performed.
Root cause (FDA determination)
Software design
Action the firm took
Siemens issued an Urgent Field Safety Notice to all customers. The letter informed customers of the potential issues and provided additional information on the implemented resolution. To resolve the problem, a system update was to be installed on each affected system under Update Instructions AX054/10/S. For questions phone (610) 448-3237.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 51 Valley Stream Pkwy, Malvern, Pennsylvania 19355
Distribution
- Distribution pattern
- USA - Nationwide Distribution
Timeline
- Recall initiated
- 2010-09-09
- Posted by FDA
- 2010-10-26
- Terminated
- 2012-09-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #94452. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.