Recalls / —
—#96439
Product
Physiological Monitoring, Patient Monitor M31250 Information Center Local Database Releases A, B, E.01, G, H, J, K, L, & M of Philips IntelliVue Information Center Software used in conjunction with PAS-210 Kit Remote Active Speakers. The intended use of the Information Center Software is to display physiologic waves, parameters and trends, formal data for strip recordings and printed reports and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.
- FDA product code
- MHX — Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K081983
- Affected lot / code info
- serial #s KVN 0001 through KVN 1071
Why it was recalled
Some PAS-201 active speaker assemblies in use with the Philips IntelliVue Information Center (PIIC) have experienced intermittent audio or loss of audio. Speaker failures may delay recognition of an alarm condition, if operators do not recognize that there is no alarm sound from the monitor and the monitor is not connected to another audible alarm annunciation system.
Root cause (FDA determination)
Device Design
Action the firm took
Philips notified all affected customers with an Urgent Medical Device Correction notification letter dated December 17, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to observe their system for any of the described symptoms. If they experience intermittent audio or there is no sound from their PAS-210 speaker while it is not connected to another audible alarm annunciation system, customers were instructed to remove the speaker from use and contact their local Philips service representative. Customers were advised not to rely exclusively on the audible alarm system for patient monitoring, the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment. ACTIONS PLANNED BY PHILIPS Philips will replace affected speaker assemblies free of charge. A Philips Healthcare representative would contact the customer when the replacement was available. Field Service Engineers will remove and replace defective speakers to correct the problem. The defective speakers will be returned to vendor (Beach Wire) to be discarded. This correction will be done free of charge for customers.
Recalling firm
- Firm
- Philips Healthcare Inc.
- Address
- 3000 Minuteman Road, Andover, Massachusetts 01810
Distribution
- Distribution pattern
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA WI, WV, WY, and Canada
Timeline
- Recall initiated
- 2010-12-17
- Posted by FDA
- 2011-01-27
- Terminated
- 2017-04-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #96439. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.