Product

FlexTray Procedure Delivery System, Product Codes: KBW71SXL- Bariatric Surgery Trays, TBW13SXL- Bariatric Surgery Trays, XBB14B - Bariatric Surgery Trays, and XBB65S - Bariatric Surgery Trays. Intended for transection, resection, and/or creation of anastomoses.

FDA product code
LRPTray, Surgical
Device class
Class 1
Medical specialty
General, Plastic Surgery
510(k) numbers
K061156
Affected lot / code info
Lot Numbers: E4MR6R, E4ML22, E4MG62, E4MJ7M, and E4MN2V.

Why it was recalled

The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint to the device shaft is insufficient.

Root cause (FDA determination)

Other

Action the firm took

Ethicon Endo-Surgery issued "Urgent Expanded Device Recall Event 1981" notification letters dated April 7, 2009, to their customers. They provided a list of all Affected Product Codes of the ETS FLEX45 Articulating Endoscopic Linear Cutters (including the Product Code: ATS45 from the 1st recall) and Affected Flex Tray Codes subject to recall. The notification document reminds the Operating Room and Materials Management customers of the problem which necessitated the previous recall (Z-0361-2011). It goes on to instruct their customers to immediately discontinue use of the affected products. They are to complete and fax back the Business Reply Form, Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle, Replacement product should be received within 30-60 days after return, Customers can contact Stericycle concerning this recall at 1-888-365-4680, reference Event 1981.

Recalling firm

Firm
Ethicon Endo-Surgery Inc
Address
4545 Creek Rd, Cincinnati, Ohio 45242-2803

Distribution

Distribution pattern
Nationwide Distribution -- USA including the following states, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MO, MS, MT, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, and Puerto Rico.

Timeline

Recall initiated
2009-03-27
Posted by FDA
2011-02-15
Terminated
2013-02-15