Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.
- FDA product code
- IWB — System, Radiation Therapy, Radionuclide
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K973441
- Affected lot / code info
- SG33620363, US68537171, SG10821625, 3713G01281, SG03321319, SG33520213, US19502892, C160, SG0320704, SG44120295, 00000000000002, SG31630678, SG24820212, SG24820194, SG44221038, SG34620630, SG40920412, SG31020878, A4986A, SG11520160, SG44720158, SG41720310, SG43820213, 3416G09539, SG24820139, US19257221, SG33820630, 19476882, SG13320399H, SG40520126, SG50425261, SG41720317, US66320311, US19490144, SG21220130, US68154038, US19501776, SW ONLY, SG32520836, 3416G08464, SG21220113, SG44925638, SG04120289, SG51220179, SG2030679, US6824174, SG22920793, SG4820232, 3628G06513, 000000000000019, 00000000000004, US19310946, 3416G05185, US69131975, SG42520530, SG23320693, US68104248, SG23720396, SG24820231, SG14720637, SG43520759, US68513855, SG44820136, SG20220701, SG44120292, US19476874, US6863578, SG04020719, 3416G07810, SG43720660, SG44925635, US68538435, US19322820, SG33320960, US19319754, SG31620624, US10102906, SG44925637, SG43520477, US69172955, SG42720416, 3711C04231, US68154724, SG0620611, SG20220192, SG33620337, US19311023, SG35020182, US19239028, SG50425266, SG51920134, 000000000000011, SG44925633, SG24021335, 0304011939, A4986A, SG34220901, SG1220108, SG32120124, US69132547, SG43720550, US19476879, 6146A06124, SG41920535, 6412A00007, US68500234, US19322145, SG20220694, US68115508, SG41620718, SG10720353, 000000000000014, SG43520485, SG42820297, 000000000000015, SG41020267, US68104599, IE24488098, SG11421969, SG41720337, US6836799, US19344670, SG11720734, US68064889, SG20220679, 000000000000016, SG20220699, SG11421978, SG34220707, US19378673, SG34420200, SG20220143, SG50425271, US10025765, US90345020, US19345857, and SG13420315.
Why it was recalled
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
Root cause (FDA determination)
Other
Action the firm took
Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.
Recalling firm
- Firm
- Elekta, Inc.
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, Georgia 30092-3011
Distribution
- Distribution pattern
- Worldwide Distribution
Timeline
- Recall initiated
- 2005-10-01
- Posted by FDA
- 2011-02-14
- Terminated
- 2011-03-02