FDA Device Recalls

Recalls /

#97584

Product

Xcelerate Specialty Patella Milling System Per File K2399; Catalog Number I-K2399PR00, I-K2399PR10; Stryker Howmedica Osteonics, Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The Xcelerate patella milling system (as detailed in surgical protocol LSPK35) was designed to prepare a patella to accept all Scorpio or Duracon patellar implants using either a resurfacing or recessed technique.

FDA product code
HTJGauge, Depth
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
All lots

Why it was recalled

Stryker Orthopaedics has become aware that the 10 mm on the bone remaining depth gage is incorrect thus allowing for the reamer to leave as little as 8.9mm (in a nominal condition) patella bone.

Root cause (FDA determination)

Device Design

Action the firm took

Stryker sent an Urgent Product Recall letter dated January 27, 2011, via FedEX for January 28, 2011, delivery with return receipt. The letter identified the product, the problem, and the action to be taken by the customer. Customers were insructed to fax back the attached Product Recall Acknowledgement Form within 5 days to 201-831-6069. Customers were also instructed to contact the hospitals in their territory that have the affected product to arrange return of the product. Product should be returned to the attention of: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey, 07430. The orange Product Remediation sticker should be attached to the return, indicating the Product Remediation # RA 2009-296. Mark the outer box with words "Product Recall." For questions regarding this recall call (201) 972-2100.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Drive, Mahwah, New Jersey 07430

Distribution

Distribution pattern
Nationwide Distribution

Timeline

Recall initiated
2011-01-27
Posted by FDA
2011-03-31
Terminated
2012-10-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #97584. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.