Recalls / —
—#98879
Product
Terumo Cardioplegia Delivery Set- Item Number: 16130 Product Usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery.
- FDA product code
- DTR — Heat-Exchanger, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K823451
- Affected lot / code info
- Lot Numbers: MN08, MNI5, MM11, MM18, MM25. Manufacturing dates October 1I, 2010 through Novemberl5, 20 I 0.
Why it was recalled
Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
Root cause (FDA determination)
Device Design
Action the firm took
Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 28 Howe St, Ashland, Massachusetts 01721-1305
Distribution
- Distribution pattern
- Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Timeline
- Recall initiated
- 2011-03-01
- Posted by FDA
- 2014-10-31
- Terminated
- 2014-11-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #98879. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.