FDA Device Recalls

Recalls /

#102859

Product

The Resolution II Clip, 155 cm size and 235 cm size, Boxes of 1, 10, and 20 clips consists of a pre-loaded, radiopaque single use clip on a flexible delivery system. Endoscopic Clipping Device, Hemorrhoidal Ligator. The device is used for clip placement within the GI tract for hemostatis ( mucosal defects, bleeding ulcer, arteries, polyps); endoscopic marking; and anchoring to affix jejunal feeding tubes to the wall of the small bowel, as a supplementary method for closure of GI tract luminal perforation <20 mm that can be treated conservatively.

FDA product code
FHNLigator, Hemorrhoidal
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K102764
Affected lot / code info
Catalog M00522500: Lot 1ML1013103, ML00000004, ML00000129, Catalog M00522501: Lot 1ML1013104, ML00000071, ML00000148, Catalog M00522502: Lot 1ML1013102, ML00000042, ML00000131, Catalog No. M00522501: Lot 1ML1011201, 1ML1013101, 1ML1020202, 1ML1021402, 1ML1022202, ML1022213,ML00000013, ML00000019, L00000058,ML00000073, ML00000098, ML00000120,  Catalog No. M00522511: Lot 1ML1010605, 1ML1012602, 1ML1013105, 1ML1020203, 1ML1021403, 1ML1022203,1ML1022214, ML00000014, ML00000020,ML00000043, ML00000059, ML00000074,ML00000099, ML00000119, Catalog No. M00522512: Lot 1ML1011905, 1ML1020201, 1ML1021401, 1ML1022201, 1ML1022215, ML00000018,ML00000032, ML00000040, ML00000072,ML00000097, ML00000135, and ML00000137

Why it was recalled

Boston Scientific has received reports where, following clip deployment, the alignment of the clip and delivery system does not allow for the withdrawal of the delivery system away from the clip, resulting in tissue trauma, increased bleeding and the clip being pulled off tissue as the delivery system is withdrawn.

Root cause (FDA determination)

Other

Action the firm took

Boston Scientific sent an URGENT MEDICAL DEVICE RECALL letter dated August 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Medical Device Recall to their customers. Customers were instructed to: Immediately discontinue use of any affected product. Complete and return the Reply Verification Tracking Form by fax to 508-683-5578. Package and return any affected product to: Boston Scientific Corporation Distribution Center Attn: QA Returns 500 Commander Shea Blvd Quincy, MA 02171 For questions regarding this recall call 508-683-5578.

Recalling firm

Firm
Boston Scientific Corporation
Address
100 Boston Scientific Way, Marlborough, Massachusetts 01752-1234

Distribution

Distribution pattern
Worldwide Distribution - including USA (nationwide)

Timeline

Recall initiated
2011-07-19
Posted by FDA
2011-08-22
Terminated
2012-06-22
Status

Source: openFDA Device Recall endpoint. Recall record ID #102859. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.