Product

Whole body Computed Tomography X-Ray System (Brilliance 64 and Ingenuity CT), Brilliance 64 M/N 728231, Ingenuity CT M/N 728326, Dist. By Philips Medical Systems, Cleveland, OH

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K033326
Affected lot / code info
System Code # 728231, Serial # 91003, 95130, 95157, 95504 & 300030; System Code # 728326, Serial # 300010 & 300018.

Why it was recalled

A number of non-conformances were identified during internal and external validation testing of the Brilliance 64 and Ingenuity CT systems with software version Beta 3 (version 4.0.0.xx368).

Root cause (FDA determination)

Software change control

Action the firm took

Philips Healthcare sent an Urgent Medical Device Correction letter dated February 17, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all memebers of their staff who need to be aware of the contents in the communication. Customers were asked to retain a copy of the communication with the equipment instruction for Use. For further information or support concerning this issue, customers should contact their local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center at 1-800-722-9377, option 5. For questions regarding this recall call 440-483-7000.

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Road, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Worldwide Distribution - USA including OH and the following countries: Australia, China, Belgium, Israel & Sweden.

Timeline

Recall initiated
2012-02-17
Posted by FDA
2012-03-23
Terminated
2013-02-22