Product

Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. Intellispace Portal is a medical software system that allows multiple users to remotely access the system from compatible computers on a network.

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K012009, K033326, K060937
Affected lot / code info
Model number: 881001; serial numbers: 80818, 85088, 81178, 82006, 82038, 81063, 82018, 79935, 81150, 81013, 81173, 85068, 79932, 79943, 80911, 82019, 81027, 85049, 85043, 82008, 81119, 81089, 85108, 85026, 85146, 80994, 81021, 81182, 81017, 85067, 81161, 81164, 79936, 85103, 81147, 85110, 85082, 81179, 81035, 85046, 81031, 81129, 81004 & 85104.

Why it was recalled

In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.

Root cause (FDA determination)

Software design

Action the firm took

Philips Healthcare sent an Urgent-Medical Device Correction letter dated May 23, 2012, via certified mail to each consignee. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the communication. The letter informed affected customers that a Philips Service Engineer will contact them for implementation of the software update on the affected systems free of charge. For any further information or support, customers can contact their local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts).

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Road, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Worldwide Distribution - US (nationwide) including the states of: AR, LA, MI, NC, NY, OH, OK and PA and the countries of: Australia, Canada, Denmark, France, Germany, Israel, Netherlands, Spain and Switzerland.

Timeline

Recall initiated
2012-05-22
Posted by FDA
2012-07-12
Terminated
2013-03-28