FDA Device Recalls

Recalls /

#125211

Product

GEMINI TF Big Bore PET/CT, Computer Tomography System, Philips Healthcare System, Cleveland, OH

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K081135
Affected lot / code info
Model #882476, Serial #'s: 9004, 9005, 9006, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9206, 9207, 9208, 9209, 9210, 9211, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, 9221, 9222, 9223, 9224, 9225, 9226, 9228, 9229, 9231.

Why it was recalled

The TF Big Bore patient table experienced mechanical binding during horizontal table motion resulting in an automatic Emergency Stop (E-stop). An E-Stop will interrupt a current scan, resulting in an incomplete study.

Root cause (FDA determination)

Software design

Action the firm took

ON 12/5/2013 the firm sent Medical Device Correction Letters to their customers.

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Rd, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Product was shipped to the following states: AR, CA, CO, FL, IL, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, PA, TX, VA, VT & WI. Product was also shipped to the following countries: Belgium, Canada, China, Denmark, France, Germany, Italy, Japan, Netherlands, Saudi Arabia & Venezuela.

Timeline

Recall initiated
2013-12-05
Posted by FDA
2014-02-12
Terminated
2015-08-25
Status

Source: openFDA Device Recall endpoint. Recall record ID #125211. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.