Recalls / —
—#138770
Product
GEMINI TF 16 Slice CT/PET System. The Philips GEMINI PET/CT Imaging Systems are a family of integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) systems suitable for a wide range of diagnostic applications.
- FDA product code
- KPS — System, Tomography, Computed, Emission
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K052640
- Affected lot / code info
- Model #882470 GEMINI TF 16 Slice (453567983931); Serial Number: 7052, 7060, 7093, 7096, 7121, 7123, 7128, 7129, 7185, 7187, 7220, 7226 & 7570.
Why it was recalled
Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Root cause (FDA determination)
Software design
Action the firm took
On 7/6/2015 the firm sent Field Safety Notices to their customers.
Recalling firm
- Firm
- Philips Medical Systems (Cleveland) Inc
- Address
- 595 Miner Rd, Cleveland, Ohio 44143-2131
Distribution
- Distribution pattern
- Worldwide Distribution - US Nationwide, Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Israel, Italy, Japan, Lebanon, Netherlands, Portugal, Saudi Arabia, South Africa, Switzerland, Thailand, Venezuela & Vietnam.
Timeline
- Recall initiated
- 2015-07-01
- Posted by FDA
- 2015-07-22
- Terminated
- 2018-06-26
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #138770. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.