Recalls / —
—#144029
Product
30" (76 cm) Appx. 3.3 ml, Admin Set w/ 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, ChemoLock, Hanger, Item No. CL3511 sterile devices intended for the infusion and withdrawal of fluids.
- FDA product code
- ONB — Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K131549
- Affected lot / code info
- 2775926 2812305 2817810 2822507 2869949 2906006 2935267 2956661 3000036 3021000 3022814 3050749 3093071 3106072 3112028 3113234 3116570 3122924 3124606 3128409 3146957 3154612 3158607 3160103 3172391 3179951 3182579
Why it was recalled
ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
A customer notification letter dated 2/25/16 was sent to all customers who purchased the IV Administration Sets with ChemoLock Port to inform them that ICU Medical has identified an issue with a connection in an isolated number of I.V. Infusion administration sets that incorporate the ChemoLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak. Customers are informed of the actions to be taken. Customers are instructed to return and complete the recall response form to ICU Medical via Fax (801) 264-1755 or by email: recall@icumed.com. Customers with questions or require assistance relating to the recall are instructed to contact Customer Service Monday through Friday between the hours of 8:30AM-4:00PM Pacific Time, (866) 829-9025 and select option 8, or email at customerservice@icumed.com.
Recalling firm
- Firm
- ICU Medical, Inc.
- Address
- 951 Calle, Amanecer San Clemente, California 92673-6212
Distribution
- Distribution pattern
- US: NV, OH, MO, CA, FL, VA, TX, NJ, CT, MI, AZ, MA, CO Non-US: DE, AE, AT, PL, HU, SE, IL, JP, JO, SA,
Timeline
- Recall initiated
- 2016-02-25
- Terminated
- 2016-08-26
- Status
- —