—#157929
Product
SQ-RX 1010 Pulse Generator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing
- FDA product code
- LWS — Implantable Cardioverter Defibrillator (Non-Crt)
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P110042
- Affected lot / code info
- All serial numbers
Why it was recalled
The device can deliver an atypical amount of energy due to memory corruption inside the device.
Root cause (FDA determination)
Software design
Action the firm took
The firm issued notifications dated June, 2017, beginning 6/29/2017 to physicians (implanting and patient follow-up physicians) via hand delivery by their sales representatives. The firm estimated approximately 30% of the U.S. physicians would be receiving the notification via hand delivery. Hand delivery by affiliates in foreign countries to customers began approximately 6/30/2017. Overnight mail letters were issued to the U.S. and foreign physicians starting on/about 7/7/2017 who did not receive hand- delivered notifications.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 4100 Hamline Ave N, Saint Paul, Minnesota 55112-5700
Distribution
- Distribution pattern
- Worldwide - US Nationwide distribution, including Puerto Rico, U.S. Virgin Island, and Guam, There was also worldwide foreign distribution, including Canada.
Timeline
- Recall initiated
- 2017-06-29
- Posted by FDA
- 2017-08-16
- Terminated
- 2019-02-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #157929. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.