Recalls / —
—#160797
Product
Philips Healthcare Brilliance iCT computed tomography x-ray system
- FDA product code
- JAK — System, X-Ray, Tomography, Computed
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K060937
- Affected lot / code info
- Brilliance iCT - 728306 Devices manufactured 03/05/2009 to present.
Why it was recalled
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
Root cause (FDA determination)
Device Design
Action the firm took
Philips Healthcare is informing consignees of this issue through a Customer Information Letter (CIL_FCO72800690). Philips Key Market representatives will distribute the letter internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order. Customers are notified that if they hear an unusual noise coming from the gantry, or scan halts, the Philips representative should be contacted as soon as possible. Philips will be inspecting the affected systems, tightening and replacing the reaction ring screws as needed at no additional cost.
Recalling firm
- Firm
- Philips Medical Systems (Cleveland) Inc
- Address
- 595 Miner Rd, Cleveland, Ohio 44143-2131
Distribution
- Distribution pattern
- 107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts
Timeline
- Recall initiated
- 2017-11-08
- Terminated
- 2020-04-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #160797. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.