FDA Device Recalls

Recalls /

#160850

Product

Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Compia MRI. Labeled as the following: a. Compia MRI CRT-D DF1 (Product No. DTMC1D1); b. Compia MRI Quad CRT-D DF4 (Product No.DTMC1QQ ) Implantable devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. Some devices can also provide resynchronization therapy, including sequential biventricular pacing (CRT-D devices).

FDA product code
NIKDefibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P010031S550, P010031S513
Affected lot / code info
a. Product No. DTMC1D1: UDI 00643169705807 (Serial No. RPW200289H);  b. Product No. DTMC1QQ: UDI 00643169565272 (Serial No. RPL201034H) (EXPANSION: Serial No. RPL200052H, RPL200534H, RPL201039H, RPL201059H, RPL201235H, RPL201236H, RPL201454H, RPL201645H, RPL201664H, RPL201234H)

Why it was recalled

Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

Root cause (FDA determination)

Process design

Action the firm took

On 12/19/2017, Medtronic verbally notified consignees and retrieved eight devices at sites that had affected inventory on hand. On 01/22/2018, Medtronic provided a recall notification letter to physicians following 48 patients implanted with affected devices and risk managers of those medical facilities, recommending that physicians strongly consider prophylactic device replacement for patients implanted with an affected device. Medtronic Technical Services is available to assist physicians with questions at 800-723-4636. Medtronic Patient Service is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Central Time). The recall was expanded on 03/15/2018 and customers were notified by letter. Affected product, not yet implanted, was retrieved on about 03/08/2018. Instructions in the physicians following patients implanted with the additional affected devices included prophylactic device replacement for those patients at higher risk, including patients whose clinical history indicates prior need for high-voltage therapy and/or for pacemaker-dependent patients. The physicians should carefully weigh the risks and benefits of device replacement. If it is determined that replacement is not warranted, instructions included the following: programming changes to reduce the potential for high-voltage charges associated with arrhythmia detection and therapies, such as enabling ATP before charging for fast ventricular rhythms or programming a separate fast VT via VF zone with ATP; Continue three-month in-clinic or remote follow-ups to verify device functionality because inability to interrogate a device or a failed remote monitoring transmission may be an indication that internal arcing has occurred; and to advice patients to seek medical attention immediately if they experience new or unexpected symptoms suspicious for a ventricular arrhythmia. Medtronic Patient Services is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Centra

Recalling firm

Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Address
8200 Coral Sea St Ne, Mounds View, Minnesota 55112-4391

Distribution

Distribution pattern
US Nationwide Distribution.

Timeline

Recall initiated
2017-12-19
Posted by FDA
2018-02-01
Terminated
2021-08-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #160850. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.