Product

Capio RP; Model #M0068321010. Intended for use in general suturing applications during open surgery to assist in the placement of suture material in tissues at the operative site.

FDA product code
PWIInstrumentation, Surgical Mesh, Urogynecologic, Transvaginal Repair Of Pelvic Organ Prolapse
Device class
Class 2
Medical specialty
Obstetrics/Gynecology
Affected lot / code info
UDI:08714729470212 Batch numbers:  17813516 18653139 19987270 18014753 18886132 20278129 18015797 19173287 20604455 18015964 19335466 20844362 18234939 19491590 21178815 18237323 19693866 21227473 21386746

Why it was recalled

Potential for Capio sutures to break and /or detach.

Root cause (FDA determination)

Device Design

Action the firm took

Boston Scientific sent Urgent Medical Device Correction letters, dated February 12, 2018, to customers via FED-EX two day delivery. Customers were informed of the updates to the affected product instructions for use and advised to read the notification carefully and immediately post the information in a visible location near the product to ensure it is easily accessible to all users of the device. Also, the notification should be passed on to any health professional of the customer's organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Customers with questions should contact their local BSC (Boston Scientific) Representative. **No product is being recalled and you are not required to return product to Boston Scientific** There is no impact to previously implanted mesh devices.

Recalling firm

Firm
Boston Scientific Corporation
Address
100 Boston Scientific Way, Marlborough, Massachusetts 01752-1234

Distribution

Distribution pattern
Worldwide Distribution -- US, Brazil, Guatemala, Ecuador, Malaysia, Czech Republic, United Kingdom, Argentina, Honduras, Peru, Singapore, Denmark, Iceland, Chile, Nicaragua, Venezuela, Japan, Egypt, Israel, Canada, Panama, Taiwan, United Arab Emirates, Finland, Ireland, Mexico, Dominican Republic, Hong Kong, Austria, France, Italy, Costa Rica, Trinidad and Tobago, Australia, Bahrain, Georgia, Islamic Republic of Iran, El Salvador, Colombia, New Zealand, Belgium, Germany, Jordan, Norway, Russia, Pakistan, Slovakia, Portugal, South Africa, Oman, Saudi Arabia, Poland, Slovenia, Qatar, Spain, Kenya, Lebanon, Mauritius, Netherlands, Sweden, Turkey, Kuwait, Malta, Namibia, Romania, Switzerland, and Greece.

Timeline

Recall initiated
2018-02-12
Posted by FDA
2018-05-04
Terminated
2020-04-10