Recalls / —
—#162764
Product
Capio RP; Model #M0068321010. Intended for use in general suturing applications during open surgery to assist in the placement of suture material in tissues at the operative site.
- FDA product code
- PWI — Instrumentation, Surgical Mesh, Urogynecologic, Transvaginal Repair Of Pelvic Organ Prolapse
- Device class
- Class 2
- Medical specialty
- Obstetrics/Gynecology
- Affected lot / code info
- UDI:08714729470212 Batch numbers: 17813516 18653139 19987270 18014753 18886132 20278129 18015797 19173287 20604455 18015964 19335466 20844362 18234939 19491590 21178815 18237323 19693866 21227473 21386746
Why it was recalled
Potential for Capio sutures to break and /or detach.
Root cause (FDA determination)
Device Design
Action the firm took
Boston Scientific sent Urgent Medical Device Correction letters, dated February 12, 2018, to customers via FED-EX two day delivery. Customers were informed of the updates to the affected product instructions for use and advised to read the notification carefully and immediately post the information in a visible location near the product to ensure it is easily accessible to all users of the device. Also, the notification should be passed on to any health professional of the customer's organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Customers with questions should contact their local BSC (Boston Scientific) Representative. **No product is being recalled and you are not required to return product to Boston Scientific** There is no impact to previously implanted mesh devices.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 100 Boston Scientific Way, Marlborough, Massachusetts 01752-1234
Distribution
- Distribution pattern
- Worldwide Distribution -- US, Brazil, Guatemala, Ecuador, Malaysia, Czech Republic, United Kingdom, Argentina, Honduras, Peru, Singapore, Denmark, Iceland, Chile, Nicaragua, Venezuela, Japan, Egypt, Israel, Canada, Panama, Taiwan, United Arab Emirates, Finland, Ireland, Mexico, Dominican Republic, Hong Kong, Austria, France, Italy, Costa Rica, Trinidad and Tobago, Australia, Bahrain, Georgia, Islamic Republic of Iran, El Salvador, Colombia, New Zealand, Belgium, Germany, Jordan, Norway, Russia, Pakistan, Slovakia, Portugal, South Africa, Oman, Saudi Arabia, Poland, Slovenia, Qatar, Spain, Kenya, Lebanon, Mauritius, Netherlands, Sweden, Turkey, Kuwait, Malta, Namibia, Romania, Switzerland, and Greece.
Timeline
- Recall initiated
- 2018-02-12
- Posted by FDA
- 2018-05-04
- Terminated
- 2020-04-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #162764. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.