Recalls / —
—#173109
Product
Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
- FDA product code
- MHX — Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K153702
- Affected lot / code info
- 866389 Patient Information Center iX B.01 (or later) 866390 Patient Information Center iX B.01 upgrade (or later)
Why it was recalled
Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values
Root cause (FDA determination)
Software design
Action the firm took
Philips initiated recall by letter on 4/29/19 identifies the reason for recall, helath risk and to follow the Action to be Taken by Customer/User section of the Customer Information Letter: Users are to be aware that the values in the patients chart or electronic medical record (EMR) may be in error and to confirm infusion pump parameters at the respective pump and only use these values to inform orders or to modify the infusion rate. Philips will provide a software update to customers using the BBraun Space LAN or Arcomed UniqueDoc infusion pumps, at no charge. For further information or support concerning this issue, contact Philips Customer Support at 1-800-722-9377. Complete the attached Reply Card and return to Philips as soon as possible
Recalling firm
- Firm
- Philips North America, LLC
- Address
- 3000 Minuteman Rd, Andover, Massachusetts 01810-1032
Distribution
- Distribution pattern
- Worldwide Distribution - US Nationwide Foreign: Andorra Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark Dominican Rep Egypt Estonia Finland France French Guiana Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kenya Korea Rep Lao Latvia Lebanon Lithuania Luxembourg Macedonia Malaysia Maldives Martinique Mauritius Mayotte Mexico Mongolia Morocco Myanmar Namibia Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestine Panama Peru Philippines Poland Portugal Puerto Rico Qatar R¿union Romania Russia Russian Fed. Saudi Arabia Serbia Singapore Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Turkey United Kingdom Uruguay Utd Arab Emir Utd.Arab.Emir. Uzbekistan Viet Nam Vietnam White Russia
Timeline
- Recall initiated
- 2019-04-29
- Terminated
- 2024-11-22
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #173109. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.