Recalls / —
—#174966
Product
Ingenuity Core128 Model # 728323, computed tomography x-ray system
- FDA product code
- JAK — System, X-Ray, Tomography, Computed
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K160743
- Affected lot / code info
- System Serial Number 6123, 30016, 30023, 300068, 30153, 32000 through 320513, 52023, 52028 336001 through 336038, 336040 through 336051, 336053, 336055 through 336061, 336063 through 336379, 336382 through 336454, 336701 through 336811, 345218, 760014, 21003-05, 320309, 32053, 32060, 336039-1 336143, 336147, 336331, 7175_06
Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Root cause (FDA determination)
Other
Action the firm took
Philips Healthcare issued a Field Safety Notification (FSN)72800720/88200522 1. Identify affected product 2. There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images. A. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when Start Final Recon is selected prior to display of the ECG wave at the top of the acquisition window. B. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs if ECG leads become disconnected mid-acquisition or when the acquisition is halted prematurely due to an application crash. C. By system design, in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase 4. Determine whether you have a potentially affected system, then determine your AFFECTED PRODUCTS software version. To identify the software version of your product: Click the Help button, Select About and the software version is then displayed, Software versions 4.x, 3.x, and 2.x are potentially affected. 5. Take the Action listed in the field safety notice. Please retain a copy with the equipment Instruction for use. If you need any further information or support concerning this issue, please contact your local Philips representative: North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377 and follow the prompts).
Recalling firm
- Firm
- Philips Medical Systems (Cleveland) Inc
- Address
- 595 Miner Rd, Cleveland, Ohio 44143-2131
Distribution
- Distribution pattern
- Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, and Yemen
Timeline
- Recall initiated
- 2019-06-24
- Posted by FDA
- 2019-08-06
- Terminated
- 2022-10-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #174966. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.