FDA Device Recalls

Recalls /

#179826

Product

OmniDiagnost Eleva System codes 708028 708027

FDA product code
OWBInterventional Fluoroscopic X-Ray System
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K032046
Affected lot / code info
Serial codes 152 49   514 179 611 C@23273 46 560 555 464 14 90 198 15 326 143*003 451 267 FR90-111 417 59 390 164 135 23 143 80 219 24 78 73 15 79 109 624 71 44 576 70 460 62 98 63 620 147 100 68 122 422 12 54 490 97 138 20 130 67 ON100031011SN273 241 435 125 89 47 17 33 134 574

Why it was recalled

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Root cause (FDA determination)

Component design/selection

Action the firm took

On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.

Recalling firm

Firm
Philips North America, LLC
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Nationwide domestic distribution. Foreign distribution worldwide.

Timeline

Recall initiated
2019-12-30
Terminated
2022-11-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #179826. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.