FDA Device Recalls

Recalls /

#182529

Product

Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

FDA product code
MWIMonitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K090785, K131382
Affected lot / code info
Product #: 989803191021 Manufacturing date codes: prior to 2020-03

Why it was recalled

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Philips Healthcare sent a customer notification letter dated June 2020 explaining the potential issues observed with their products. This letter instructs the consumer to contact the company with any reports of issues associated with this failure.

Recalling firm

Firm
Philips North America, LLC
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.

Timeline

Recall initiated
2020-06-18
Terminated
2023-01-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #182529. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.