FDA Device Recalls

Recalls /

#184355

Product

31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3020; ChemoLock Port w/Spiros, Red Cap, REF: CL-35; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger, REF: CL3951; 40" (102 cm) Appx 7.7 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Bag Hanger, REF: CL3962; 31" (79 cm) Appx 3.5 ml, Admin Set w/ChemoLock Bag Spike with Additive Port, 20 Drop In-Line Chamber w/15 Micron Filter, Spiros, Purple Cap, Hanger, REF: CL3999

FDA product code
ONBClosed Antineoplastic And Hazardous Drug Reconstitution And Transfer System
Device class
Class 2
Medical specialty
General Hospital
Affected lot / code info
Lots: 4867937, 4751234, 4799262, 4894789, 4787230, 4867947, 4897075, 4720480, 4735684, 4758456, 4775068, 4794504, 4860043, 4799897, 4787234, 4808045, 4860044                                       ,

Why it was recalled

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Root cause (FDA determination)

Process control

Action the firm took

On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or ProductComplaintsPP@icumed.com; Customer Service, 1-866-829-9025, option 8 or, customerservice@icumed.com, (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.

Recalling firm

Firm
ICU Medical, Inc.
Address
951 Calle, Amanecer San Clemente, California 92673-6212

Distribution

Distribution pattern
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Timeline

Recall initiated
2020-08-31
Status

Source: openFDA Device Recall endpoint. Recall record ID #184355. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.