FDA Device Recalls

Recalls /

#184517

Product

IcePearl" 2.1 CX L 90 Cryoablation Needle

FDA product code
GEHUnit, Cryosurgical, Accessories
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K162599
Affected lot / code info
UPN: FPRPR3617  UDIs: 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072, 7290104831072  Batch numbers: U1310, U1309, U1306, U1305, X0828, U0727, U0728, U0641, U0640, U0587, U0586, U0585, U0084, A7113, A6977, A6963, A6962, A6681, A6682, A1657, A1522, A1178, A1177

Why it was recalled

Complaint trend regarding needle shaft gas leaks.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Customers will be sent a product advisory via overnight mail.

Recalling firm

Firm
Boston Scientific Corporation
Address
1 Scimed Pl, Maple Grove, Minnesota 55311-1565

Distribution

Distribution pattern
Nationwide

Timeline

Recall initiated
2020-11-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #184517. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: IcePearl" 2.1 CX L 90 Cryoablation Needle · FDA Device Recalls