Recalls / —
—#184754
Product
Medtronic Cobalt HF CRT-D MRI SureScan DTPB2D1, MR Conditional implantable cardioverter defibrillator with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (DDE-DDDR)
- FDA product code
- NIK — Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P010031
- Affected lot / code info
- Serial Numbers: RTP601047S, RTP601055S
Why it was recalled
Under rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
Root cause (FDA determination)
Process control
Action the firm took
During the week of 15-Nov-2020, Medtronic initiated retrieval activities for 6 devices (5 US/1 OUS) for consignees in possession of potentially affected, non-implanted devices according to Medtronic records. Consignees were either verbally directed by Medtronic Field Representatives to quarantine product, or in the case of consigned units, the product was retrieved from consigned inventory and returned to Medtronic for engineering evaluation. The firm will follow with a letter to be disseminated on 12/04/2020 to the risk managers.
Recalling firm
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Address
- 8200 Coral Sea St Ne, Mounds View, Minnesota 55112-4391
Distribution
- Distribution pattern
- Distribution to US states of GA, PA, NJ, and France
Timeline
- Recall initiated
- 2020-11-15
- Terminated
- 2025-12-23
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #184754. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.