Product

Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

FDA product code
LLZSystem, Image Processing, Radiological
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K042880, K061029
Affected lot / code info
Model# 882310,882311,882313 All Lots

Why it was recalled

A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.

Root cause (FDA determination)

Device Design

Action the firm took

On 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.

Recalling firm

Firm
Philips North America, LLC
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latvia, Libya, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Kingdom, Venezuela.

Timeline

Recall initiated
2020-11-19
Terminated
2023-04-19