Product
Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- FDA product code
- LLZ — System, Image Processing, Radiological
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K042880, K061029
- Affected lot / code info
- Model# 882310,882311,882313 All Lots
Why it was recalled
A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
Root cause (FDA determination)
Device Design
Action the firm took
On 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.
Recalling firm
- Firm
- Philips North America, LLC
- Address
- 3000 Minuteman Rd, Andover, Massachusetts 01810-1032
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latvia, Libya, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Kingdom, Venezuela.
Timeline
- Recall initiated
- 2020-11-19
- Terminated
- 2023-04-19
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #184817. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.