Recalls / —
—#186466
Product
CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)
- FDA product code
- NKE — Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P890003S386
- Affected lot / code info
- all serial numbers
Why it was recalled
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
Root cause (FDA determination)
Software design
Action the firm took
Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2-day UPS delivery of the consignee letter. Confirmation delivery of the consignee letter will be obtained. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. Consignees were advised that Medtronic representatives can assist in installing the software update on programmers. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software. A courtesy notification will be provided to physicians. No confirmation will be required for the courtesy notification.
Recalling firm
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Address
- 8200 Coral Sea St Ne, Mounds View, Minnesota 55112-4391
Distribution
- Distribution pattern
- Worldwide
Timeline
- Recall initiated
- 2021-04-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #186466. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.