Product

Ingenia Elition S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781357

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K173451
Affected lot / code info
Serial Number Unique Device Identifier (UDI#) Model: 781357 46029 (01)00884838088108(21)46029 46053 (01)00884838088108(21)46053 46045 (01)00884838088108(21)46045 46027 (01)00884838088108(21)46027 46042 (01)00884838088108(21)46042 46084 (01)00884838088108(21)46084 46086 (01)00884838088108(21)46086 46083 (01)00884838088108(21)46083 46030 (01)00884838088108(21)46030 46031 (01)00884838088108(21)46031 46050 (01)00884838088108(21)46050 46049 (01)00884838088108(21)46049 46052 (01)00884838088108(21)46052 46076 (01)00884838088108(21)46076 46085 (01)00884838088108(21)46085 46024 (01)00884838088108(21)46024 46044 N/A 46061 N/A 46001 (01)00884838088108(21)46001 46005 (01)00884838088108(21)46005 46011 (01)00884838088108(21)46011 46012 (01)00884838088108(21)46012 46057 (01)00884838088108(21)46057 46048 (01)00884838088108(21)46048 46021 (01)00884838088108(21)46021 46066 (01)00884838088108(21)46066 46078 (01)00884838088108(21)46078 46081 (01)00884838088108(21)46081 46060 (01)00884838088108(21)46060 46004 (01)00884838088108(21)46004 46018 (01)00884838088108(21)46018 46020 (01)00884838088108(21)46020 46013 (01)00884838088108(21)46013 46015 (01)00884838088108(21)46015 46017 (01)00884838088108(21)46017 46016 (01)00884838088108(21)46016 46023 (01)00884838088108(21)46023 46033 (01)00884838088108(21)46033 46034 (01)00884838088108(21)46034 46037 (01)00884838088108(21)46037 46025 (01)00884838088108(21)46025 46041 (01)00884838088108(21)46041 46026 (01)00884838088108(21)46026 46032 (01)00884838088108(21)46032 46036 (01)00884838088108(21)46036 46059 (01)00884838088108(21)46059 46039 (01)00884838088108(21)46039 46075 (01)00884838088108(21)46075 46046 (01)00884838088108(21)46046 46068 (01)00884838088108(21)46068 46058 (01)00884838088108(21)46058 46054 (01)00884838088108(21)46054 46082 (01)00884838088108(21)46082 46069 (01)00884838088108(21)46069 46003 (01)00884838088108(21)46003 46019 (01)00884838088108(21)46019 46008 (01)00884838088108(21)46008 46007 (01)00884838088108(21)46007 46067 (01)00884838088108(21)46067 46089 N/A 46002 (01)00884838088108(21)46002 46014 (01)00884838088108(21)46014 46065 (01)00884838088108(21)46065 46077 (01)00884838088108(21)46077

Why it was recalled

When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment

Root cause (FDA determination)

Device Design

Action the firm took

Philips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is detected in the HDU, the Philips FSE will replace the necessary mechanical component(s) to resolve the issue. If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377) and reference FCO78100527.

Recalling firm

Firm
Philips North America Llc
Address
222 Jacobs St, Cambridge, Massachusetts 02141-2289

Distribution

Distribution pattern
Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.

Timeline

Recall initiated
2021-08-11
Status

Source: openFDA Device Recall endpoint. Recall record ID #188802. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.