Recalls / —
—#210463
Product
Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100
- FDA product code
- OAE — Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P220032
- Affected lot / code info
- GTIN 08714729992660
Why it was recalled
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
Root cause (FDA determination)
Device Design
Action the firm took
Boston Scientific disseminated a letter to Physicians/Healthcare Professionals on 10/10/2024 via courier. The letter notified its consignees of updates to the instructions for use (IFU) for the POLARx and POLARx FIT Cryoablation Catheter related to the risk of atrioesophageal (AE) fistula. The firm requested that the information be shared with clinicians who use the use the Boston Scientific POLARx Cryoablation System, including the POLARx Catheter, the POLARx FIT Catheter and the SMARTFREEZE Console, and to share the communication with any other organization to which these devices may have been transferred. Users were directed to immediately post the information on or near the product to ensure this information is easily accessible to all users of the device.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 4100 Hamline Ave N, Saint Paul, Minnesota 55112-5700
Distribution
- Distribution pattern
- Worldwide
Timeline
- Recall initiated
- 2024-10-10
- Posted by FDA
- 2024-11-12
- Status
- —