Recalls / —
—#215650
Product
Preventive Maintenance (PM) Kit ICEFX, Material Number (UPN) SPRPM8000
- FDA product code
- GEH — Unit, Cryosurgical, Accessories
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- 510(k) numbers
- K234002
- Affected lot / code info
- Batch: 108159110
Why it was recalled
Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
Root cause (FDA determination)
Process control
Action the firm took
An Urgent Medical Device Correction dated 8/18/25 was sent to customers. Customers should cease use of the systems listed in Table 1 and segregate them until a Boston Scientific representative has corrected this issue. A Boston Scientific Field Service Engineer will contact your facility to schedule the correction of your system. For patients in whom an impacted device has already been used, there are no specific recommendations beyond the local standard post-procedure care. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative. The Acknowledgement Form enclosed with this product correction must be completed and returned even if you do not have any affected units. An Urgent Medical Device Removal notification letter dated 8/18/25 was sent to a customer regarding the PM Kit ICEFX. Instructions: 1. Further distribution or use of any devices (Table 1) affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page three. 5. Return affected devices. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 1 Scimed Pl, Maple Grove, Minnesota 55311-1565
Distribution
- Distribution pattern
- Worldwide - US Nationwide and the countries of Canada, France, Germany and Italy. (Preventive Maintenance Kit distributed to Canada only.)
Timeline
- Recall initiated
- 2025-08-18
- Posted by FDA
- 2025-09-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #215650. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.