FDA Device Recalls

Recalls /

#32088

Product

VAXCEL MINI-STICK COAXIAL DILATOR SET. Each kit contains a radiopaque Coaxial Dilator (5 Fr), a 21 Ga. echogenic entry needle, and a .018 in. / 0,46 mm guidewire with a floppy lip. Cat. #45-994. Firm on the label: Boston Scientific.

FDA product code
DREDilator, Vessel, For Percutaneous Catheterization
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
910482, Exp. 10/31/05

Why it was recalled

Kits containing 5 Fr dilators may be labeled as 4 Fr. Kits containing 4 Fr dilators may be labeled as 5 Fr.

Root cause (FDA determination)

Other

Action the firm took

Letter dated 3/5/04 with instructions to return product.

Recalling firm

Firm
Boston Scientific Corporation
Address
1 Boston Scientific Pl, Natick, Massachusetts 01760-1536

Distribution

Distribution pattern
42 domestic hospitals nationwide and one distributor in Canada.

Timeline

Recall initiated
2004-03-05
Posted by FDA
2004-07-20
Terminated
2005-10-26
Status

Source: openFDA Device Recall endpoint. Recall record ID #32088. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: VAXCEL MINI-STICK COAXIAL DILATOR SET. Each kit contains a radiopaque Coaxial Dilator (5 Fr), a 21 · FDA Device Recalls