FDA Device Recalls

Recalls /

#40593

Product

Vaxcel Chest Ports - Standard/T/PU/9/1.8/2.8/63 Catalog Number : 45-315

FDA product code
KOBCatheter, Suprapubic (And Accessories)
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K030083
Affected lot / code info
Lot Number: 1005248

Why it was recalled

The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.

Root cause (FDA determination)

Other

Action the firm took

Boston Scientific intiated the recall on 7/22/05 to direct accounts and to distributors on 7/ 27 /05. Distributors are requested to conduct sub recalls to the medical facility level. A revised letter dated 8/2/05 was sent to 18 US accounts for one lot (#1020895) omitted in the letter of 7/22/05.

Recalling firm

Firm
Boston Scientific Corporation
Address
1 Boston Scientific Pl, Natick, Massachusetts 01760-1536

Distribution

Distribution pattern
Nationwide VA Hopsitals Foreign: Canada, Africa, Japan, Israel, Japan, Nertherlands, Venezula

Timeline

Recall initiated
2005-07-22
Posted by FDA
2005-09-02
Terminated
2007-06-20
Status

Source: openFDA Device Recall endpoint. Recall record ID #40593. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.