FDA Device Recalls

Recalls /

#64547

Product

Philips Medical Systems, GEMINI TF 16 & 64 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems, Philips Medical Systems, Cleveland, OH 44143

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K051170, K052640
Affected lot / code info
Gemini TF   Product Numbers/Manufacturer Numbers: TF 64 slice: 4535 679 94741, TF 16 slice: 4535 679 83931.  Catalog Numbers: TF 64 slice: 882471, TF 16 slice: 882470.   Serial Numbers: 7001-7058, 7062, 7064, and 7066.

Why it was recalled

Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.

Root cause (FDA determination)

Software design

Action the firm took

Philips Medical notified domestic customers via Product Safety Notification dated May 30, 2007. The product safety notification includes a Technical Bulletin notifying customers of the issue and how to avoid it. This information will be used until the software correction becomes available. The software correction will be released in October 2008. Once software is available, the firm will begin correcting the field.

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Road, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Worldwide Distribution - USA, Australia, Belgium, Canada, China, Denmark, England, Finland, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Netherlands, Poland, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Taiwan, Turkey, and Venezuela.

Timeline

Recall initiated
2007-09-21
Posted by FDA
2008-01-31
Terminated
2012-03-02
Status

Source: openFDA Device Recall endpoint. Recall record ID #64547. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.