FDA Device Recalls

Recalls /

#66436

Product

Axis Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K964712
Affected lot / code info
Model Numbers: 210714, 210880 and 211037 Part Numbers: 4535 665 13461, or N210714 Axis System, 3/8" Crystals 4535 665 16601, or N210880 Axis System 3/4" Crystals 4535 665 12921, or N211037 Axis System 3/8 Precision 4535 665 13501, or N211039 Axis System 3/4 Precision 4535 679 46301, or N211280 Axis System 3/8 Precision 4535 671 67421, or N211282 Axis System 3/4 Precision (Note: Due to a part number scheme change at the firm, either number is representative of the identified system)

Why it was recalled

Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.

Root cause (FDA determination)

Component design/selection

Action the firm took

Consignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Road, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Worldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Netherlands, Norway, Philipines, Poland, Singapore, South Africa, Spain, Switzerland, and United Kingdom.

Timeline

Recall initiated
2007-11-21
Posted by FDA
2008-09-19
Terminated
2012-02-25
Status