Product

Rotate Motion Shunt Resistor Kit, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K964712
Affected lot / code info
Lot Numbers: 4535 664 01241 and N753360

Why it was recalled

Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.

Root cause (FDA determination)

Component design/selection

Action the firm took

Consignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.

Recalling firm

Firm
Philips Medical Systems (Cleveland) Inc
Address
595 Miner Road, Cleveland, Ohio 44143-2131

Distribution

Distribution pattern
Worldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Netherlands, Norway, Philipines, Poland, Singapore, South Africa, Spain, Switzerland, and United Kingdom.

Timeline

Recall initiated
2007-11-21
Posted by FDA
2008-09-19
Terminated
2012-02-25
Status