iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
- FDA product code
- IYO — System, Imaging, Pulsed Echo, Ultrasonic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K072517
- Affected lot / code info
- UPNs: H74900036130, H749FG000940, H749ILAB120INS0, H749ILAB120INSR0, H749ILAB220INS0, H749240INS0, M004EPIL120INS0. Serial numbers 0000004082, 0000004385, 0000004386, 0000004387, 0000004388, 0000004389, 0000004390, 0000004391, 0000004392, 0000004393, 0000004396, 0000004397, 0000004398, 0000004399, 0000004416, 0000004417, 0000004418, 0000004419, 0000004420, 0000004421, 0000004422, 0000004423, 0000004424, 0000004425, 0000004426, 0000004427, 0000004428, 0000004429, 0000004430, 0000004431, 0000004432, 0000004433, 0000004434, 0000004435, 0000004436, 0000004439, 0000004441, 0000004442, 0000004443, 0000004444, 0000004445, 0000004446, 0000004447, 0000004448, 0000004453, 0000004454, 0000004455, 0000004456, 0000004468, 0000004469, 0000004472, 0000004473, 0000004480, 0000004487, 0000004488, 0000004489, 0000004490, 0000004491, 0000004494, 0000004495, 0000004496, 0000004497, 0000004498, 0000004505, 0000004506, 0000004507, 0000004512, 0000004513, 0000004514, 0000004515, 0000004516, 0000004520, 0000004521, 0000004522, 0000004523, 0000004524, 0000004525, 0000004526, 0000004529, 0000004538, 0000004539, 0000004548, 0000004549, 0000004550, 0000004551, 0000004556, 0000004557, 0000004558, 0000004559, 0000004564, 0000004565, 0000004566, 0000004567, 0000004572, 0000004573, 0000004574, 0000004575, 0000004580, 0000004581, 0000004582, 0000004583, 0000004584, 0000004585, 0000004586, 0000004587, 0000004610, 0000004702, 0000004703, 0000004704, 0000004705, 0000004717, 0000004718, 0000004719, 0000004720, 0000004726, 0000004728, 0000004729, 0000004730, 0000004731, 0000004732, 0000004737, 0000004738, 0000004751, 0000035228, 0000035229, 0000035390, 0000035391, 0000036314, 0000094718, 0009430893, 0009503458, 0009543381, 0009565345, 0009902558, 0009602559, 0009626567, 0009640091, 0009828054, 0009828055, 0009828344, 0011008866.
Why it was recalled
Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.
Root cause (FDA determination)
Process control
Action the firm took
On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 47900 Bayside Pkwy, Fremont, California 94538-6515
Distribution
- Distribution pattern
- Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.