Product

iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538

FDA product code
IYOSystem, Imaging, Pulsed Echo, Ultrasonic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K072517
Affected lot / code info
UPNs: H74900036130, H749FG000940, H749ILAB120INS0, H749ILAB120INSR0, H749ILAB220INS0, H749240INS0, M004EPIL120INS0. Serial numbers 0000004082, 0000004385, 0000004386, 0000004387, 0000004388, 0000004389, 0000004390, 0000004391, 0000004392, 0000004393, 0000004396, 0000004397, 0000004398, 0000004399, 0000004416, 0000004417, 0000004418, 0000004419, 0000004420, 0000004421, 0000004422, 0000004423, 0000004424, 0000004425, 0000004426, 0000004427, 0000004428, 0000004429, 0000004430, 0000004431, 0000004432, 0000004433, 0000004434, 0000004435, 0000004436, 0000004439, 0000004441, 0000004442, 0000004443, 0000004444, 0000004445, 0000004446, 0000004447, 0000004448, 0000004453, 0000004454, 0000004455, 0000004456, 0000004468, 0000004469, 0000004472, 0000004473, 0000004480, 0000004487, 0000004488, 0000004489, 0000004490, 0000004491, 0000004494, 0000004495, 0000004496, 0000004497, 0000004498, 0000004505, 0000004506, 0000004507, 0000004512, 0000004513, 0000004514, 0000004515, 0000004516, 0000004520, 0000004521, 0000004522, 0000004523, 0000004524, 0000004525, 0000004526, 0000004529, 0000004538, 0000004539, 0000004548, 0000004549, 0000004550, 0000004551, 0000004556, 0000004557, 0000004558, 0000004559, 0000004564, 0000004565, 0000004566, 0000004567, 0000004572, 0000004573, 0000004574, 0000004575, 0000004580, 0000004581, 0000004582, 0000004583, 0000004584, 0000004585, 0000004586, 0000004587, 0000004610, 0000004702, 0000004703, 0000004704, 0000004705, 0000004717, 0000004718, 0000004719, 0000004720, 0000004726, 0000004728, 0000004729, 0000004730, 0000004731, 0000004732, 0000004737, 0000004738, 0000004751, 0000035228, 0000035229, 0000035390, 0000035391, 0000036314, 0000094718, 0009430893, 0009503458, 0009543381, 0009565345, 0009902558, 0009602559, 0009626567, 0009640091, 0009828054, 0009828055, 0009828344, 0011008866.

Why it was recalled

Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.

Root cause (FDA determination)

Process control

Action the firm took

On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.

Recalling firm

Firm
Boston Scientific Corporation
Address
47900 Bayside Pkwy, Fremont, California 94538-6515

Distribution

Distribution pattern
Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.

Timeline

Recall initiated
2000-04-11
Posted by FDA
2008-09-23
Terminated
2009-04-06
Status