FDA Device Recalls

Recalls /

#74734

Product

Boston Scientific Easy Core" Biopsy System, 18 ga. x 21 cm, REF 43-454, UPN M001434541, Sterilized with ethylene oxide gas. UPN for inner pouch M001434540. Indicated for use endoscopically or percutaneously to retrieve tissue sampling of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.

FDA product code
FCGBiopsy Needle
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K040893
Affected lot / code info
11520676 and 11551300

Why it was recalled

Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.

Root cause (FDA determination)

Device Design

Action the firm took

Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated October 21, 2008. The letter was addressed to The Hospital Field Action Contact. The letter described the problem and products being recalled. The letter stated that it was very important that the attached product table be used carefully and consider both inner and outer packaging product codes when searching for affected/recalled product as the UPN numbers on the inner and outer labeling are different. Requested to discontinue use of product, identify, and segregate and return affected products to BSC. Letter also requested to complete and return the Reply Verification Tracking Form.

Recalling firm

Firm
Boston Scientific Corporation
Address
1 Scimed Pl, Maple Grove, Minnesota 55311-1565

Distribution

Distribution pattern
Nationwide Distribution --- including states of AL, AZ, AR, CA, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO,MT, NE , NV, NH, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and PR.

Timeline

Recall initiated
2008-10-07
Posted by FDA
2008-11-28
Terminated
2011-12-10
Status