Recalls / —
—#74738
Product
Boston Scientific Easy Core" Biopsy System, 18 ga. x 15 cm, REF 43-459, UPN M001434591, Sterilized with ethylene oxide gas. UPN for inner pouch M001434590. Indicated for use endoscopically or percutaneously to retrieve tissue sampling of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.
- FDA product code
- FCG — Biopsy Needle
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K040893
- Affected lot / code info
- 11518343, 11535832, and 11634771
Why it was recalled
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
Root cause (FDA determination)
Device Design
Action the firm took
Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated October 21, 2008. The letter was addressed to The Hospital Field Action Contact. The letter described the problem and products being recalled. The letter stated that it was very important that the attached product table be used carefully and consider both inner and outer packaging product codes when searching for affected/recalled product as the UPN numbers on the inner and outer labeling are different. Requested to discontinue use of product, identify, and segregate and return affected products to BSC. Letter also requested to complete and return the Reply Verification Tracking Form.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 1 Scimed Pl, Maple Grove, Minnesota 55311-1565
Distribution
- Distribution pattern
- Nationwide Distribution --- including states of AL, AZ, AR, CA, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO,MT, NE , NV, NH, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and PR.
Timeline
- Recall initiated
- 2008-10-07
- Posted by FDA
- 2008-11-28
- Terminated
- 2011-12-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #74738. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.