FDA Device Recalls

Recalls /

#80888

Product

Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN# DESCRIPTION M0064102020 Percuflex All Purpose Drainage Catheters wit Fader Tip and Locking Pigtail 8 Fr M0064102030 Percuflex All Purpose Drainage Catheters with Fader Tip and Locking Pigtail 10 Fr Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scintific Corporation, One Boscton Scintific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080. Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.

FDA product code
FGEStents, Drains And Dilators For The Biliary Ducts
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
All codes   ¿

Why it was recalled

Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.

Root cause (FDA determination)

Labeling False and Misleading

Action the firm took

Consignees were sent a Boston Scientific " Urgent Medical Device Field Correction Notification" letter dated March 18, 2009. The letter was addressed to "The Hospital Field Action Contact". The letter described the problem, product affected and the revised catheter removal instructions. They also requested the return of the Acknowledgement Form.

Recalling firm

Firm
Boston Scientific Corporation
Address
1 Scimed Pl, Maple Grove, Minnesota 55311-1565

Distribution

Distribution pattern
Worldwide Distribution -- USA including DC, Puerto Rico and Guam and countries of United Arab Emirates Austria, Bosnia and Herzegovina, Bahrain, Switzerland, Germany, Egypt, Spain, Finland, France, Great Britain, Ireland, Italy, Kuwait, Netherland, Pakistan, Portugal, Saudi Arabia, Sweden, Slovak Republic, Tunisia, Turkey, South Africa, Netherlands, Costa Rica, Japan, Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Ecuador, El Salvador, Guatemala, Honduras, Hong Kong, Macau, Malaysia, Mexico New Zealand, Panama, Peru, Philippines, Puerto Rico, Singapore, South Korea, Sri Lanka, Taiwan, Thailand, Uruguay, and Venezuela.

Timeline

Recall initiated
2009-03-18
Posted by FDA
2009-05-21
Terminated
2012-02-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #80888. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.