Product

Ultraflex Tracheobronchial Stent System. Uncovered - 18/80 Catalog Number: M00569280. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.

FDA product code
JCTProsthesis, Tracheal, Expandable
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K012883
Affected lot / code info
Lot Numbers: 11021753 11021758 11480991 11480993 11721839 11722425 11859270 118641921

Why it was recalled

Suture related device failures during deployment and repositioning of the stent

Root cause (FDA determination)

Device Design

Action the firm took

Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.

Recalling firm

Firm
Boston Scientific Corporation
Address
100 Boston Scientific Way, Marlborough, Massachusetts 01752-1234

Distribution

Distribution pattern
Worldwide Distribution - Only US distributed product is affected by recall

Timeline

Recall initiated
2010-05-17
Posted by FDA
2010-06-25
Terminated
2011-10-20