Recalls / —
—#91906
Product
Ultraflex Tracheobronchial Stent System. Uncovered - 20/80 Catalog Number: M00569310. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
- FDA product code
- JCT — Prosthesis, Tracheal, Expandable
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- 510(k) numbers
- K012883
- Affected lot / code info
- Lot Numbers: 9820027 9820415 11092859 11093048 11752551 11962126 11962127
Why it was recalled
Suture related device failures during deployment and repositioning of the stent
Root cause (FDA determination)
Device Design
Action the firm took
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Recalling firm
- Firm
- Boston Scientific Corporation
- Address
- 100 Boston Scientific Way, Marlborough, Massachusetts 01752-1234
Distribution
- Distribution pattern
- Worldwide Distribution - Only US distributed product is affected by recall
Timeline
- Recall initiated
- 2010-05-17
- Posted by FDA
- 2010-06-25
- Terminated
- 2011-10-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #91906. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.